Eppendorf 5430R Main Frequency Converter PCB- Troubleshooting
- Janet Ferrao
- Sep 22
- 1 min read
Updated: Oct 16
Version 1


Board Connections for Version 2

New models come with Input filters — eliminate the filters and run Mains from your Power on Switch directly to the Main board- please observe connections.
Sahara Simba Electronics
Troubleshooting
Service Functions
Service Function | Description of Function |
1 | Display test and key test |
2 | Test lid function |
3 | Test rotor detection / rotor impulses |
4 | Determine acceleration and deceleration times |
5 | Initialize EEPROM |
6 | Test overspeed cut-off |
7 | Endurance test program |
8 | Error messages and data of last 4 runs |
9 | Operating hours/load cycles |
10 | Test USB interface |
11 | Test individual aggregates (fan/compressor) |
12 | Run with adjustable constant compressor frequency |
13 | Check imbalance detection |
14 | Machine test (for plant inspections only) |
To Enter the Service Mode:
Switch off the centrifuge.
Press the ▲ and ▼ keys simultaneously and switch on the mains power switch while doing so.
Navigation through Service Mode:
Use the ▲ and ▼ keys to select the service function.
Use the menu key or the ► arrow key to start the service function.