top of page
Search

Eppendorf 5910R/RI

Now servicing the 5910R/RI Main Frequency Converter board Please inquire for pricing. The Display will be needed to repair the 5910R PCB.

5810R 5804R Rotor Recognition Test

Please perform the following for Rotor Recognition. Go to Service Mode 8. ( To enter the Service Mode - hold up and down keys while...

Eppendorf 5427R no E-FAN error

Remove fan and test with external power supply. If plastic casing is broken or cracked use epoxy and align the fan so it can freely...

Eppendorf 5415D Board Disassembly

Remove Rotor Remove 4 Phillips screws on the bottom of the Centrifuge housing. Lift housing to the side as shown below . Disconnect...

Eppendorf 5424R Initialize

To “initialize” the unit you have to enter the Service mode Hold RPM key down Turn the Main Switch ON SE 1 will appear on your display...

Eppendorf 5424R Main FC PCB connections

Connect Imbalance/Temperature, Fans, Condensor Sensor, Rotor Speed Sensor Next connect S800 (Ribbon Cable) from small PCB Communication...

Scams— Please be aware!!!

I received a request for a quote to purchase 10 Eppendorf centrifuges . The scammers sent me a fictitious Purchase Order … looks...

Eppendorf 5415R Error 20

Temperature sensor error displayed as soon as you turn on the centrifuge. This error is related to the sensor inside the rotor chamber....

Eppendorf Mastercycler

The best cycler Eppendorf has made is the Mastercycler Gradient. The epGradient is a flawed design with many failures. And now a way to...

bottom of page