Service Mode #6
Push the lid down and then press START
Push “temp “ button to turn the compressor on
CAUTION : THE COMPRESSOR WILL OVERHEAT IF YOU RUN IT WITHOUT RUNNING THE COMPRESSOR FAN.
To turn on the compressor fan push the “time” button
The Up and Down arrow keys are used to control the speed of the fan
Push the Program key to test the Blower fan
Push thr Speed key to test the electronic fans on the Frequency Converter PCB